FEI quata 3D: OMNI probe problem fixed on July 27

OMNI probe was bent and the setting was messed up during my vacation. After returning to Irvine, I replaced the bent tip with a new one, but could not solve the setting problem. With the on-line assistance of FEI engineer, the problem could not be solved either. FEI engineer will be on campus on July 25.
Please report problems when they occur. Users are not expected to fix any machine problems.

Fischione ion mill down on June 23 and fixed on Sept 26.

The ion mill was down again. Testing results indicate thatĀ it needs to be shipped to vendor for repair. It is uploaded in a crate and ready to go. The PO request was approved. Calit2 business office is working with Fischione to have it shipped to the vendor now. The repair could take 4 weeks or longer. Please read comments for further updates.

Fischione Ion Mill down on May 26 and up on June 17

An ion mill problem was reported yesterday. This problem was found to be associated with the low gun which was used heavily recently. I have contacted the vendor andĀ received a quote today. A service request has been submitted today. Waiting for approval now. The instument could be down for two weeks or longer because of PO and gun service.

CM-20 TEM: Sample holder and condenser aperture

It was reported that the double tilt specimen holder did not work. I found that the cable was damaged by our user. User should hold the metal part to disconnect the cable from the holder. User should never pull the cable to disconnect it. We will try to fix the cable, or order a new cable.
The condenser aperture is adjusted. It works well now.
TEM EDS is conditioned and calibrated. It work well now. I may want to condition it again in near future. Please check the comments in the instrument status.
Thanks